Introduction
And Description
The Digital Haemoglobin (Hb) Reader is designed,
keeping in the view to make the analysis
simpler. The cumbersome step of calculation
is deleted while using the digital Hb Reader.
The final result of the unknown test will
be displayed directly in conc. (gm/dl).
There is no need of calculations. However
O.D. Switch is also given in case of optical
density is required to crosscheck the reading.
A micro controller has been incorporated
in the system for accuracy and reproducibility
of findings.
The Digital Hb Reader is
a compact, elegant, very reliable micro
controller based equipment to produce accurate
results. An optical glass filter of 540
± 10 nanometers and square cuvette
have been incorporated to give accurate
results. This Hb Reader has been designed
to operate in the following two modes :
-
A. STD.MODE [Concentration of 15 gm/dl has
been fed]
B. NON STD. MODE [Absorbance of 0.30 (O.D.)
has been fed]
In case, the CONC. or ABS.
of STD. are different from 15 gm/dl or 0.30
respectively, the user can feed the required
values with the help of switch no. 3 &
4 (given on the KEYPAD on front panel).
| Specifications |
| Wavelength Filter |
- 540 ± 10 nm optical
glass filter |
| Minimum Volume |
- 1 ml |
| Display |
- 7 segment, Red LED, 3.5 digit |
| Indicator |
- 5 mm LED Indicators provided for
different modes. |
| Lamp |
- 6 V / 10W Tungsten Lamp |
| Detector |
- Si photodiode |
| Input Supply |
|
| Mains Operation |
- 230V ± 10%, 50Hz, AC |
| Battery Operation |
- 12V |
| Dimension |
- 12" x 11.5" x 6.25" |
| Weight |
- 5 kg Approx. |
| |
|
| Standard Accessories |
|
| Lamp |
- 6V / 10 W - 2 Nos. |
| Fuse |
- 350 mA - 2 Nos. |
| Cuvette |
- Square - 5 Nos. |
| Dust Cover |
- Plastic - 1 No. |
| Instruction Manual |
- Book Type - 1 No. |
| Battery Lead |
- 1 Pair |
Installation Requirement
(A) Voltage 230V ± 10 AC, 50Hz (or
AC stabilizer)
(B) Earthing must be proper.
(C) Room should be dust free.
(D) Vibration should be avoided (electrical
or mechanical)
(E) Radiation should be avoided.
(F) Humidity should be less than 60-70%.
Unpack the equipment and plug into three-pin
socket having proper earth. Leave it for
5-10 minutes to warm up before use.
|