Introduction and Description
The Digital Haemoglobin (Hb/Bl) Reader is designed, keeping in view to make the analysis simpler. The cumbersome step of calculation is deleted while using the digital Hb/Bl Reader. The final result of the unknown test will be displayed directly in Conc. There is no need of calculations. O.D. can also be measured. However ‘AUTO ZERO’ facility is also provided.

The digital Hb/Bl Reader with LCD alpha-numeric display is a compact, elegant, light weight and reliable micro controller based equipment to produce accurate results.

An optical glass LED (546nm) and square cuvette has been incorporated to give accurate results. This instrument has been designed to operate in the following modes :-

A. MEASUREMENT MODE
B. CALIBRATION MODE
C. PHOTO COL – 546nm MODE

Specification

Wavelength - 546nm– a green colour LED of 546nm
Minimum Volume - 1 ml
Display - LCD Alpha Numeric
Light Source - 546nm – A green colour LED
Detector - Si Photo Diode
Input Supply  
Mains Operation - 230 V + 10%, 50 Hz, AC
Battery Operation - 12 V
Dimension - 10.25 x 9 x 5.5”
Weight - 1.80 Kgs
   
Standard Accessories
Fuse - 350 mA - 2 Nos
Cuvette - Square - 5 Nos
Dust Cover - Plastic - 1 No.
Instruction Manual - Book Type - 1 No.
Battery Lead - 1 Pair


Installation Requirement
a.) Voltage 230 V + 10 AC, 50 Hz (or AC stabilizer).
b.) Earthing must be proper.
c.) Room should be dust free.
d.) Vibration should be avoided (Electrical or Mechanical).
e.) Radiation should be avoided.
f.) Humidity should be less than 60-70%.
Unpack the equipment and plug into three-pin socket having proper earth. Leave it for 5-10 minutes to warm up before use.

 
Operating Modes
» Calibration Mode
In this mode, you can change the FED value of CONC/FACTORS of Standards or Calibrator with the help of the Switch 3 and 4

» Measurement Mode
In this mode, if the CONC/FACTORS of Standard solutions matched with the FED values, then there is no need of any calibration. You can directly go to the testing.

» Photo Col. Mode
In this mode, the instrument will act as a colorimeter at 546 nm. It gives only the absorbance of the Standard or Tests.

Operating Steps
1. Switch ‘ON’ the instrument by the switch given at the back of the machine.
- Display will show KM 003 Hb/Bl / Haemoglobin Mode
2. Press K1 Display will show, BILIRUBIN MODE
3. Again Press K1 Display will show PHOTO COL. – 546



Items of dopplers/ PH Meters Category
Doppler Hand Held (Model no. KDM – 001)
Ultrasound Doppler Vascular (Model no. KDV – 05)
Ultrasonic Foetal Heart Monitor
Micro Controlled Digital Foetal Doppler (Model no KM – 002)
Hb-Bl Reader (Model no. KM – 003)
Digital pH/mV Meter
Digital pH Meter, Hand Held
Digital Hb Reader (BMS)
Digital Colony Counter
Digital TDS/Cond. Meter, Hand Held
Needle and Syringe Destroyer.
Ultra Sonic Digital Dopplers.

 PRODUCT CATEGORIES
Anaesthesia Products
Analyzers
Autoclave Portable Vertical-Horizontal
Baby Incubator
Body Care Aids
Colorimeters
Dopplers/ PH Meters
Examination Light
General Items
General Surgical Instruments
Gynecological Products
Hospital Garments
I.V. Cannulas
Laboratory Equipment
Microscopes
Misc. Medical Products
Orthopaedic Implants
Stethoscopes
Suction Units, Spares & Vacuum Extractor